Integrated unit for clinical accuracy
Two ways of easy flushing- Controlled squeeze & Snap-tab
Available with flushing of 3cc/hr, 30cc/hr or without flushing
Trouble free priming for easy Zeroing
Closed system to minimize Transducer Cross contamination
|
 |
Sensitivity
Zero Balance
Input Impedance - OHMS (with reusable cable)
Output Impedance - OHMS
Calibration Requirement
Operating Temperature Range
Storage Temperature Range
Low flow rate (3ml/hr varient)
Low flow rate (30ml/hr varient)
High flow rate
Operating Pressure
Overpressure
Overpressure, (with flush & specific dp/dt)
Symmetry |
5.0 ± 1% µ v/v/mmHg
0 ± 40 (mmHg)
350 Nominal
300 Nominal (mmHg)
100 ± 3 (mmHg)
+15 to +40 °C
25 to +70 °C
1.9 ml <flow< 5.0 ml / hr @ 276 mmHg
20 ml <flow< 40 ml / hr @ 276 mmHg
minimum flow of (125ml / min) @ 276 mmHg
30 to 300 (mmHg)
-400 to 6000 (mmHg)
>_ 10,000** (mmHg)
1 ± 5% |
B-KT-164 (Standard Single Kit)
|
 |
| B-KT-165 (Standard Double Kit) |
 |
| B-KT-190 (Standard Triple Kit) |
 |
| Customized kits available as per OR/ICU Protocol |